Access 2 Reaction Vessels
Brand(s)
Last updated
Summary
Product
Access 2 Reaction Vessels
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| Access 2 Reaction Vessels | P243971 | 81901 |
| Access 2 Reaction Vessels | P244345 | 81901 |
| Access 2 Reaction Vessels | P250005 | 81901 |
| Access 2 Reaction Vessels | P250001 | 81901 |
| Access 2 Reaction Vessels | P243943 | 81901 |
| Access 2 Reaction Vessels | P244202 | 81901 |
| Access 2 Reaction Vessels | P244344 | 81901 |
| Access 2 Reaction Vessels | P250002 | 81901 |
| Access 2 Reaction Vessels | P250004 | 81901 |
| Access 2 Reaction Vessels | P250003 | 81901 |
Issue
Instrument testing replicated failures seen in the field (e.g., pipetting, rake, wash carousel errors) for suspect RV lots. Inspections confirmed high percentages of rigid gate vestige exceeding specifications across multiple lots.
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
Beckman Coulter, Inc.
250 S. Kraemer Blvd., Brea, California, United States, 92821
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-81215
Get notified