Health product recall

Aisys Cs2

Last updated

Summary

Product
Aisys Cs2
Issue
Medical devices - Performance issue
What to do

Continue using anesthesia system if no cracks. If crack observed, contact manufacturer. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalog number
Aisys Cs2 More than 10 numbers, contact manufacturer. 1011-9050-000

Issue

The affected anesthesia systems base can have a crack in a specific location in the rear of the system. If excessive load is applied to the device, a cracked base could potentially fracture, resulting in a tip or overbalance of the anesthesia system, which can result in potential injury. 

Recall start date: June 3, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies

Datex-Ohmeda, Inc.

3030 Ohmeda Drive, Madison, Wisconsin, United States, 53707-7550

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-64266

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe