CARDIOSAVE Intra-Aortic Balloon Pump (IABP) (2018-05-08)
- Starting date:
- May 8, 2018
- Posting date:
- May 25, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-66886
Affected products
CARDIOSAVE Intra-Aortic Balloon Pump (IABP)
Reason
Maquet/Getinge has received complaints involving the CARDIOSAVE IABPs regarding ingress of fluids into the IABP affecting various electronic circuit boards. This situation would prevent initiation or continuation of therapy.
Affected products
CARDIOSAVE Intra-Aortic Balloon Pump (IABP)
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
0998-00-0800-XX
Companies
- Manufacturer
-
Datascope Corp.
1300 Macarthur Blvd.
Mahwah
07430
UNITED STATES