Health product recall

Extension Table Top

Brand(s)
Last updated

Summary

Product
Extension Table Top
Issue
Medical devices - Performance issue
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Extension Table Top

5

1140.20F0

Issue

For commercially released 1140.20f0 operating table tops it has been determined that a coincidence of 1. Lacking maintenance, 2. Not following the IFU in terms of lever orientation to prevent collision with the extension bars, and 3. Inattention to if at least one side is locked properly may lead to a loss of supporting functionality.

Recall start date: October 24, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies

Maquet GmbH

Kehler Strasse 31, Rastatt, Germany, 76437

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-72142