Gentec Medical Air Regulators - 287 MA Series (2018-06-27)
- Starting date:
- June 27, 2018
- Posting date:
- July 13, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-67222
Affected products
Gentec Medical Air Regulators - 287 MA Series
Reason
At the outlet of the Regulator, a DISS Oxygen fitting might be mistakenly installed instead of a DISS Air fitting. Users will not be able to connect Air hose to the outlet. In addition, there is wording "OXYGEN" printed on gauge face. This may mislead users that the regulators supply Oxygen gas.
Affected products
Gentec Medical Air Regulators - 287 MA Series
Lot or serial number
All lots.
Model or catalog number
GEN-287MA-25LY1
GEN-287MA-25LY2
Companies
- Manufacturer
-
GENSTAR TECHNOLOGIES CO., INC.
4525 Edison Ave
Chino
91710
California
UNITED STATES