Liberator Stationary Liquid Oxygen Device
Brand(s)
Last updated
Summary
Product
Liberator Stationary Liquid Oxygen Device
Issue
Medical devices - Performance issue
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalog number |
---|---|---|
Liberator Stationary Liquid Oxygen Device | CBB3022320074 CBB3022310852 CBB3022310843 CBB3022310827 CBB3022310823 CBB3022310842 CBB3022310845 CBB3022320077 CBB3022310792 CBB3022310844 | 13261699 13256195 |
Issue
There were inconsistencies in weld penetration on the longitudinal weld seam on the inner bottle. This can lead to premature inner vessel failure. The failure mode would be potential loss of vacuum, which could cause either an over-delivery of oxygen or the outer vessel to frost. Increased flow of oxygen from the relief valves could lead to an increased fire risk.
Recall start date: October 11, 2022
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies
Caire Inc.
2200 Airport Industrial Drive, Ball Ground, Georgia, United States, 30107
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-72172