Progressa (Frame)
Brand(s)
Last updated
Summary
Product
Progressa (Frame)
Issue
Medical devices - Performance issue
What to do
Contact the manufacturer if you require additional information.
Affected products
Affected products | Lot or serial number | Model or catalog number |
---|---|---|
Progressa (Frame) | X206AW8012 X159AW7380 X206AW8010 X206AW8011 X159AW7378 X206AW8020 X159AW7375 | P7500A000158 P7500A000842 |
Issue
During verification testing, it was discovered that a subset of the beds under test were experiencing bent compression links when raising the head section and when maximum patient weight was simulated. This failure mode was due to a change that was made to how the compression links attach to the bed frame. This may cause the compression links to bend under normal use with heavy (500 lbs) patients.
Recall start date: October 25, 2022
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Orthopaedics
Companies
Hill-Rom, Inc.
1069 State Route 46 East, Batesville, Indiana, United States, 47006-9167
Published by
Health Canada
Audience
General public
Healthcare
Industry
Recall class
Type II
Identification number
RA-72179
Get notified