Health product recall

Siemens Healthcare Magnetom

Last updated

Summary

Product
Siemens Healthcare Magnetom
Issue
Medical devices - Performance issue
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalogue number
Magnetom Aera More than 10 numbers, contact manufacturer. 10432914
Magnetom Skyra 45166 45087 145249 45934 145034 45961 45186 45370 145335 10432915
Magnetom Prisma 166247 166026 66108 10849582
Magnetom Prisma Fit 167096 167104 10849583

Issue

The service pack NXA30A-SP02 is a bug fix update to improve stability and performance of NXVA30A-based MR systems. The service pack includes various internal improvements regarding the robustness of the basic syngo and Numaris functionality. This includes:
- latest OSand internal security updates and
- higher SW upgrade, boot and service stability.
also it addresses improvements in:
- MR View & Go,
- Dot engines,
- GSP, and general protocol handling and MR measurement.
Customer made protocols and WIPs, as well as IDEA package, are not affected.

Recall start date: June 28, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies

Siemens Healthcare Gmbh

Henkestr. 127, Erlangen, Germany, 91052

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-64352

Get notified

Receive notifications for new and updated recalls and alerts by category.