TechnegasPlus Generator (2019-06-20)
- Starting date:
- June 20, 2019
- Posting date:
- July 15, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-70415
Last updated: 2019-07-15
Affected Products
TechnegasPlus Generator
Reason
On May 24th, 2019, while performing conducted emissions certification testing on the technegasplus generator to CISPR 11, the TechnegasPlus Generator recorded a measurement above the upper limit for class a conducted emissions. Specifically, the out-of-specification reading occurred during the 17 second burn cycle on the active line testing while assessing quasi-peak levels.
Affected products
TechnegasPlus Generator
Lot or serial number
More than 100 numbers, contact manufacturer.
Model or catalog number
25000
Companies
- Manufacturer
-
CYCLOMEDICA AUSTRALIA PTY LTD
UNIT 4, 1 THE CRESCENT
KINGSGROVE
2208
AUSTRALIA