Health product recall

0180/OMNERA 400T (2021-05-17)

Starting date:
May 17, 2021
Posting date:
June 17, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-75811



Last updated:
2021-06-17

Affected Products 

0180/OMNERA 400T

Reason

Cleaning of the maneuver handle or the maneuver display with excessive amount of disinfectants containing certain components pose a risk of causing a short circuit due to ingress of liquid, which in turn could cause uncontrolled up- or down movement of the overhead tube crane (OTC).

Affected products

0180/OMNERA 400T

Lot or serial number

All lots. 

Model or catalog number

180

Companies
Manufacturer

ARCOMA AB

ANNAVAGEN 1

VAXJO