1788 4K Camera
Last updated
Summary
Product
1788 4K Camera
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| 1788 4K Camera Head, C-Mount With Advanced Imaging Modality | 23H519774 | 1788210105 |
| 1788 4K Camera Control Unit With Advanced Imaging Modality | Not applicable | 1788010000 |
| 1788 4K Camera Head With Integrated Coupler With Advanced Imaging Modality | More than 10 numbers, contact manufacturer. | 1788610122 |
Issue
Stryker Endoscopy has received an increase in complaints for 1788 out of box failures due to bent connection pins. The bent pin may also damage the mating connector in the CCU. Stryker has investigated and identified a service repair issue which contributed to the bent pins noted with these devices.
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Health products - Medical devices - Gastroenterology and urology
Companies
| Stryker Endoscopy, A Div. Of Stryker Corp. Dba Stryker Endoscopy |
| 5900 Optical Court, San Jose, California, United States, 95138 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-82442
Get notified