Access hsTnI
Brand(s)
Last updated
Summary
Product
Access hsTnI
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
---|---|---|
Access hsTnI |
All lots. |
B52699 |
Issue
In remote circumstances, if a sample containing high cardiac troponin (CTNI) >55,000 pg/ml is tested using an access assay other than Access hsTnI, CTNI carryover may occur if the next test performed, using the same probe on that instrument, is Access hsTnI on a different sample.
Recall start date: November 18, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies
Beckman Coulter, Inc. |
250 S. Kraemer Blvd., Brea, California, United States, 92821 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76641