Access Sensitive Estradiol Calibrators (2020-04-22)
- Starting date:
- April 22, 2020
- Posting date:
- May 8, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-73033
Last updated: 2020-05-08
Affected Products
- Access Sensitive Estradiol Calibrators
Reason
Beckman Coulter has determined that four current access sensitive estradiol calibrator (P/N B84494) lots share the same lot numbers as four expired access free T3 calibrator (P/N A13430) lots. The access system prevents the configuration of a calibrator lot with duplicate lot numbers, which is already configured on the system.
Affected products
A. Access Sensitive Estradiol Calibrators
Lot or serial number
- 989909
- 989910
- 989911
- 989912
Model or catalog number
- B84494
Companies
- Manufacturer
-
Beckman Coulter, Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES