Acl Top Cts
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected Products |
Lot or Serial Number |
Model or catalog number |
---|---|---|
Acl Top 500 Cts |
All lots. |
00000280040, 00000280040R |
Acl Top 300 Cts |
All lots. |
00000280060R, 00000280060 |
Acl Top 550 Cts |
All lots. |
00000280045 |
Acl Top 350 Cts |
All lots. |
00000280065 |
Issue
We have become aware of a potential carryover issue that may cause elevated quality control and sample results for heparin, apixaban or rivaroxaban when hemosil liquid anti-xa reagent is used in the same run with both hemosil liquid antithrombin and hemosil q.f.a. thrombin (bovine).
Recall start date: Feb 22, 2022
Additional information
Details
Instrumentation Laboratory Co.
180 Hartwell Road, Bedford, Massachusetts
United States, 01730-2443