Acuson Sequoia Diagnostic Ultrasound System
Brand(s)
Last updated
Summary
Product
Acuson Sequoia Diagnostic Ultrasound System
Issue
Medical devices - Inaccurate test or measurement results
What to do
Contact the manufacturer if you require additional information.
Affected products
| Affected Products | Lot or serial number | Model or catalog number |
|---|---|---|
| Acuson Sequoia Diagnostic Ultrasound System | 800343 | 11148775 |
Issue
| System overestimates the circumference when the manual trace function is used. the trace circumference value is overestimated; however, the area value is accurate. the issue is present in Sequoia's VA20A/VA25 onward as the VA20A introduced a new calculation design that did not include an optimization step of discarding pixel points that are too close together, causing the close pixel points into the accumulated circumference/distance value makes the final value highly sensitive to small trackball motions. |
Recall start date: 2021-12-16
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies
| Siemens Medical Solutions Usa, Inc. |
| 22010 Se 51st Street, Issaquah, Washington, United States, 98029 |
Published by
Health Canada
Audience
General public
Health professionals
Hospitals
Recall class
Type II
Identification number
RA-63779