Health product recall

Affinity 4 Birthing Bed

Brand(s)
Last updated

Summary

Product
Affinity 4 Birthing Bed
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Affinity 4 Birthing Bed

Z347AA9101
Z346AA9094
Z347AA9102
Z346AA9095

P3700E100444
P3700E100436

Issue

Baxter Corporation is issuing an urgent medical device correction for specific Affinity 4 Birthing Beds with hilo and foot actuators due to the potential for improper crimps on the signal wires of the actuators, which could result in a loss of ability to lower the bed or the foot section of the bed.

Recall start date: May 7, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
Hill-Rom, Inc.
1069 State Route 46 East, Batesville, Indiana, United States, 47006-9167
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77477

Get notified

Receive notifications for new and updated recalls and alerts by category.