Health product recall

Alinity C & I Processing Module (2019-09-19)

Starting date:
September 19, 2019
Posting date:
November 15, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-71627



Last updated: 2019-11-15

Produits touchés

A. Alinity C Processing Module

B. Alinity I Processing Module

Reason

The development of environmental stress on the bulk solution level sensor can cause cracks. Cracks can allowair to enter the fluid line, which may result either a failure to dispense or an incomplete dispense of the bulk solution on Alinity i and Alinity c.

Affected products

A. Alinity C Processing Module

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • 03R67-01
Companies
Manufacturer

Abbott GmbH and Co. K.G.

Max Planck Ring 2

Wiesbaden

65205

GERMANY