Alinity C system Optics Assembly (2020-05-20)
- Starting date:
- May 20, 2020
- Posting date:
- May 29, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-73209
Last updated: 2020-05-29
Affected Products
Alinity C system Optics Assembly
Reason
This Product Correction letter is to inform you of an issue with the Alinity c system Optics Assembly (part number C-35016278-01). Abbott has identified the potential for a nonlinear optics response due to an issue with the Alinity c optics. The issue is with the diffraction grating within the optics assembly and can result in a lower than expected absorbance value.
Affected products
Alinity C system Optics Assembly
Lot or serial number
All lots.
Model or catalog number
03R67-01