Health product recall

Allura Xper Fd10 & Fd20 products

Last updated

Summary

Product
Allura Xper Fd10 & Fd20 products
Issue
Medical devices - Mechanical defect
What to do

Contact the manufacturer if you require additional information.

Affected products

Affected Products

Lot or serial number

Model or catalog number

Allura Xper Fd10/10 (Product Of Alluraclarity Family)

Not applicable

722027

Allura Xper Fd10 (Product Of Alluraclarity Family)

Not applicable

722026

Allura Xper Fd10

Not applicable

001443

Allura Xper Fd20 (Product Of Alluraclarity Family)

Not applicable

722028

Issue

Philips has discovered a problem with the flat detector cooling system. Due to a leak in the flat detector cooling system, cooling liquid may leak outside the drip tray of the laird chiller. If this occurs, the liquid may drip into electrical components in the R-cabinet located in the technical room which could cause damage to the system.

Recall start date: Nov 22, 2021

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies
Philips Medical Systems Nederland B.V.

Veenpluis 6, Best

Netherlands, 5684 PC

Published by
Health Canada
Audience
General public
Health professionals
Hospitals
Recall class
Type II
Identification number
RA-63712

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe