Allura Xper Systems
Last updated
Summary
Product
Allura Xper Systems
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalogue number |
---|---|---|
Allura Xper FD20 (Product Of AlluraClarity Family) | Not applicable. | 722028 |
Allura Xper FD 20 Or Table (Product Of AlluraClarity Family) | Not applicable. | 722035 |
Allura Xper FD10/10 (Product Of AlluraClarity Family) | Not applicable. | 722027 |
Allura Xper FD10 (Product Of AlluraClarity Family) | Not applicable. | 722026 |
Allura Xper FD20/20 (Product Of AlluraClarity Family) | Not applicable. | 722038 |
Allura Xper FD10 | Not applicable. | 001443 |
Allura Xper FD20/15 | Not applicable. | 722058 |
Issue
Philips has identified that the BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the Allura System will be halted and the Allura System will not start. No user messages are shown before the battery is low on power or depleted.
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies
Philips Medical Systems Nederland B.V.
Veenpluis 6, Best, Netherlands, 5684 PC
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-78125