Approach CTO-6 and Hydro ST Microwire Guide With Hydrophylic Coating
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
Approach Hydro ST Microwire Guide With Hydrophylic Coating |
15894225 |
HMW-14-190-ST |
Approach CTO-6 Microwire Guide |
15835853 |
CMW-14-300-18G |
Issue
Devices from the affected device lots have labels that state the incorrect expiration dates. Preliminary investigation into this issue identified a calculation error occurred in Cook's labeling software, which uses a calculator to determine the correct expiration date. Due to this error, the calculator assigned the maximum potential shelf life to lots in scope rather than the actual shelf life.
Recall Start Date: October 8, 2024
Additional information
Details
Cook, Incorporated
750 Daniels Way, Bloomington, Indiana, United States, 47404
Get notified
Receive notifications for new and updated recalls and alerts by category.