ARTIS icono ceiling

Last updated

Summary

Product
ARTIS icono ceiling
Issue
Medical devices - Device compatibility
What to do

Contact the manufacturer if you require additional information.

Affected products

Affected products Lot or serial number Model or catalogue number
ARTIS icono ceiling Not applicable. 11328100

Issue

The affected part is the Display ceiling suspension (DCS Fix XL) that holds the monitor in place. Over time, some screws that keep the monitor attached may slowly become loose. This happens near the x-ray indicator light. As the screws loosen gradually, a small gap can be seen underneath the indicator light. This gap increases slowly and can be noticed over a longer period of time.
If operation continues while the screws are loosening, internal cables routed within the arm may become damaged. This damage can result in the loss of the displayed image. If the condition is not detected in time or is ignored, additional screws may progressively loosen over time.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies

Siemens Healthcare Gmbh

Henkestr. 127, Erlangen, Germany, 91052

Published by
Health Canada
Audience
General public
Healthcare
Industry
Recall class
Type II
Recall date
Identification number
RA-82216