ARTIS Icono Ceiling, Floor and Biplane Systems
Brand(s)
Last updated
Summary
Product
ARTIS Icono Ceiling, Floor and Biplane Systems
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Affected products
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| ARTIS Icono Ceiling | Software version VE40A & VE40B | 11328100 |
| ARTIS Icono Floor | Software version VE40A & VE40B | 11327700 |
| ARTIS Icono Biplane | Software version VE40A & VE40B | 11327600 |
Issue
In very rare situations, after increasing the detector lift height to its maximum or minimum, the system may lose its movement capability. A reboot may not reestablish movement control. This failure can be resolved only by qualified service personnel.
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies
Siemens Healthcare GmbH
Henkestr. 127, Erlangen, Germany, 91052
Published by
Health Canada
Audience
General public
Healthcare
Industry
Recall class
Type II
Recall date
Identification number
RA-82159