Artis Zeego Multi-Axis, Artis Zee Floor, Artis ceiling, Artis zee biplane (2018-04-27)
- Starting date:
- April 27, 2018
- Posting date:
- May 11, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-66738
Affected products
- Artis Zeego Multi-Axis
- Artis Zee Floor
- Artis ceiling
- Artis zee biplane
Reason
Siemens is providing safety information for Artis zee /Q/Q.zen systems with large display DSC 5608-DC starting with serial number 1550 onwards (with l04 panel). The large display does not show an image due to a technical problem. This might occur after the large display returns from power save mode. The large display stays black without showing an error message. X-ray is still possible. The problem occurs sporadically and only when the display is returning from power save mode. It does not occur during an ongoing procedure.
Affected products
A. Artis Zeego Multi-Axis
Lot or serial number
160846
Model or catalog number
10280959
Companies
- Manufacturer
-
Siemens Healthcare GmbH
Henkestr. 127
Erlangen
91052
GERMANY
B. Artis Zee Floor
Lot or serial number
- 136556
- 136972
Model or catalog number
10094135
Companies
- Manufacturer
-
Siemens Healthcare GmbH
Henkestr. 127
Erlangen
91052
GERMANY
C. Artis ceiling
Lot or serial number
- 147519
- 147576
- 147598
- 147631
- 147632
Model or catalog number
10094137
Companies
- Manufacturer
-
Siemens Healthcare GmbH
Henkestr. 127
Erlangen
91052
GERMANY
D. Artis zee biplane
Lot or serial number
153859
Model or catalog number
10094141
Companies
- Manufacturer
-
Siemens Healthcare GmbH
Henkestr. 127
Erlangen
91052
GERMANY