Product
Asp6025 and Asp300 S
Issue
Medical devices - Performance issue
What to do

Unplug network cable. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Asp6025

SN 0712

ASP6025

Asp300 S

SN 2849

ASP300S

Issue

Leica Biosystems recognized through post market surveillance 7 vulnerabilities affecting remote care, The remote access solution produced by Axeda, (collectively known as access:7). Exploiting these vulnerabilities, attackers with network access to a target device could remotely execute code, access its file system, or alter system configurations.

Recall start date: July 14, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Pathology
Companies

Leica Biosystems Nussloch GmbH

Heidelberger Street 17-19, Nussloch, Germany, 69226

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-64396