Health product recall

Atellica CH Reaction Ring Cuvette Segment

Last updated

Summary

Product
Atellica CH Reaction Ring Cuvette Segment
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Affected products

Affected products

Lot or serial number

Model or catalog number

Atellica CH Reaction Ring Cuvette Segment

Starting with “N15” or “15"

11099326

Issue

Siemens Healthineers has confirmed, through the internal investigation of customer complaints that there is a potential for imprecision with seven (7) assays identified to have optical sensitivity when using a subset of lots of Atellica® CH Reaction Ring Cuvette segments on some Atellica CH analyzers.

Recall Start Date: October 28, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Ophthalmology
Companies

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave, Tarrytown, New York, United States, 10591

Published by
Health Canada
Audience
General public
Healthcare
Industry
Recall class
Type II
Identification number
RA-76440

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe