Health product recall

Autoimmune EIA Anti-Tissue Transglutaminase (tTG) IgA (2019-04-10)

Starting date:
April 10, 2019
Posting date:
May 10, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-69866



Last updated: 2019-05-10

Affected Products

Autoimmune EIA Anti-Tissue Transglutaminase (tTG) IgA

Reason

Elevated lot-to-lot bias were observed by comparing a new EIA anti-Q15tTG IgA autoimmune lot (catalog# 4252260, lot# 18200a) to a previous lot (catalog# 4252260, lot# 18010a). This observation has been confirmed by recent tests by Bio-Rad using internal test samples.

Affected products

Autoimmune EIA Anti-Tissue Transglutaminase (tTG) IgA

Lot or serial number

18200A

Model or catalog number

4252260

Companies
Manufacturer

Bio-Rad Laboratories, Inc.

4000 Alfred Nobel Dr.

Hercules

94547

California

UNITED STATES