Health product recall

AXIOM Vertix and AXIOM Multix

Brand(s)
Last updated

Summary

Product
AXIOM Vertix and AXIOM Multix
Issue
Medical devices - Performance issue
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

AXIOM Vertix MD Trauma

8908290

8908290

AXIOM Multix Mt

8395381
8395415
8395399

44 99 302

Issue

Siemens Healthineers was informed by the monitor supplier that 5 out of 27.000 delivered power supplies of a touch-display power supply unit were affected by overheating due to a short circuit. In one of the five events, the connected monitor caught fire. No injuries were associated with these events. To ensure the continued safe operation of your system and to safeguard its lifetime, the power supply of the touch display will be replaced preventively.

Recall start date: May 29, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies

Siemens AG

Wittelsbacherplatz 2, Muenchen, Germany, 80333

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75669

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe