Health product recall

Azurion 7 Systems

Last updated

Summary

Product
Azurion 7 Systems
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Azurion 7 B12

128

722 067

Azurion 7 B12

151

722 067

Azurion 7 M12

119

722 223

Azurion 7 M12

205

722 078

Azurion 7 M12

444

722 078

Azurion 7 M12

218

722 078

Azurion 7 M12

206

722 078

Azurion 7 M12

204

722 078

Azurion 7 M20

3459

722 224

Azurion 7 M20

874

722 224

Azurion 7 M20

2667

722 224

Azurion 7 M20

495

722 224

Azurion 7 M20

1156

722 224

Azurion 7 M20

3596

722 224

Azurion 7 M20

875

722 224

Azurion 7 M20

284

722 079

Azurion 7 M20

333

722 079

Azurion 7 M20

1374

722 079

Azurion 7 M20

783

722 079

Azurion 7 M20

750

722 079

Azurion 7 M20

679

722 079

Azurion 7 M20

352

722 079

Issue

Philips Azurion systems use solid-state drives (SSDs) in various PCs to store the operating system and image data and to support system functionality. Philips has identified that, during certain repair or service actions, SSDs that are not specified in the medical device file may have been installed. These SSDs may have different performance characteristics, which may include reduced reliability over time.

When an SSD does not perform as intended, the symptoms depend on the PC in which the SSD is installed and the function being used. this may result in loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies

Philips Medical Systems Nederland B.V.

Veenpluis 6, Best, Netherlands, 5684 PC

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-82841

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe