Health product recall

Barrel Clamp Guide (2018-04-04)

Starting date:
April 4, 2018
Posting date:
July 19, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67254

Affected products

Barrel Clamp Guide

Reason

Smith’s Medical has initiated a voluntary recall for certain Barrel Clamp Guides (part # G6000716) used in the repair of Medfusion Syringe pumps. Barrel Clamp Guides were individually sold to consignees for facility self-repair and service.

Certain Barrel Clamp Guides may have contained a ridge in the molded part. As a result, spring slippage may potentially occur resulting in the inability of the pump to recognize a syringe or the pump may misidentify the size of syringe loaded.

Affected products

Barrel Clamp Guide

Lot or serial number

P0239588

P0321444

Model or catalog number

G6000716

Companies
Manufacturer

Smiths Medical ASD, Inc.

1265 Grey Fox Road

St. Paul

55112

Minnesota

UNITED STATES