Barrel Clamp Guide (2018-04-04)
- Starting date:
- April 4, 2018
- Posting date:
- July 19, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-67254
Affected products
Barrel Clamp Guide
Reason
Smith’s Medical has initiated a voluntary recall for certain Barrel Clamp Guides (part # G6000716) used in the repair of Medfusion Syringe pumps. Barrel Clamp Guides were individually sold to consignees for facility self-repair and service.
Certain Barrel Clamp Guides may have contained a ridge in the molded part. As a result, spring slippage may potentially occur resulting in the inability of the pump to recognize a syringe or the pump may misidentify the size of syringe loaded.
Affected products
Barrel Clamp Guide
Lot or serial number
P0239588
P0321444
Model or catalog number
G6000716
Companies
- Manufacturer
-
Smiths Medical ASD, Inc.
1265 Grey Fox Road
St. Paul
55112
Minnesota
UNITED STATES