BD BACTEC MGIT 960 PZA Kit And Susceptibility Test Medium
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products | Lot or serial number | Model or catalogue number |
---|---|---|
BD BACTEC MGIT 960 PZA Kit And Susceptibility Test Medium | More than 10 numbers, contact manufacturer. | 245128 |
Issue
Through internal BD complaint trending and subsequent internal raw material testing it was identified that the BD BACTEC™ MGIT™ 960 PZA Kits, listed in attachment 1 may intermittently produce falsely resistant results for pyrazinamide (PZA) during susceptibility testing of mycobacterium tuberculosis isolates.
Recall start date: July 18, 2024
Additional information
Details
Becton Dickinson And Company (Bd) |
7 Loveton Circle, Sparks, Maryland, United States, 21152 |