Health product recall

BD Phoenix System - NMIC 1 and NMIC 93 (2018-01-09)

Starting date:
January 9, 2018
Posting date:
February 1, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-65880

Affected products

  1. BD PHOENIX SYSTEM - NMIC 1
  2. BD PHOENIX NMIC-93

Reason

There is an increased rate of very major errors (vme) for Colistin with Colistin resistant strains of the Enterobacteriaceae (17.5%). This percentage is derived from a recently-concluded external study, and likely involves emerging hetero-resistant strains..

Affected products

A. BD PHOENIX SYSTEM - NMIC 1

Lot or serial number
  • 7095829
  • 7276978
Model or catalog number

448063

Companies
Manufacturer

Becton Dickinson and Company

7 Loveton Circle

Sparks

21152

Maryland

UNITED STATES


B. BD PHOENIX NMIC-93

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • 448764
  • 448788
Companies
Manufacturer

Becton Dickinson and Company

7 Loveton Circle

Sparks

21152

Maryland

UNITED STATES