Beckman Coulter AU Immunochemistry System - AU480 Analyzer
Brand(s)
Last updated
Summary
Product
Beckman Coulter AU Immunochemistry System - AU480 Analyzer
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| Beckman Coulter AU Immunochemistry System - AU480 Analyzer | 2021104332 | AU480 |
| Beckman Coulter AU Immunochemistry System - AU480 Analyzer | 2021124493 | AU480 |
| Beckman Coulter AU Immunochemistry System - AU480 Analyzer | 2022014627 | AU480 |
| Beckman Coulter AU Immunochemistry System - AU480 Analyzer | 2021124487 | AU480 |
| Beckman Coulter AU Immunochemistry System - AU480 Analyzer | 2022014611 | AU480 |
Issue
Beckman Coulter has determined that a software defect in AU480 chemistry analyzers using Microsoft® SQL server 2022 express may cause a DB connection error that renders alarmDB unreadable and prevents instrument startup. The issue is associated with transaction log auto-growth, and may cause delays in reporting patient test results to physicians.
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies
| Beckman Coulter, Inc. |
| 250 S. Kraemer Blvd., Brea, California, United States, 92821 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-82443
Get notified
Receive emails about new and updated recall and safety alerts.