Beckman Coulter DxC 700 AU Clinical Chemistry Analyzer (2018-06-12)
- Starting date:
- June 12, 2018
- Posting date:
- October 19, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-67962
Affected products
Beckman Coulter DxC 700 AU Clinical Chemistry Analyzer
Reason
A software limitation occurs with software version 1.07 when using the analyzer maintenance help, the ise maintenance help or the consumables help icons. The issue may cause an incorrectly displayed dialog box or white screen. In some cases the system has to be rebooted. This could lead to a delay of reporting patient results.
Affected products
Beckman Coulter DxC 700 AU Clinical Chemistry Analyzer
Lot or serial number
Software version 1.07
Model or catalog number
- B86443
- B86444
- B86445
- B86446
Companies
- Manufacturer
-
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
UNITED STATES