Health product recall

Blueprint®

Brand(s)
Last updated

Summary

Product
Blueprint®
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Blueprint®

Version 4.3.1

Not applicable.

Blueprint®

Version 4.2.2

Not applicable.

Issue

Stryker has identified the following software bugs with Blueprint® versions 4.2.2 and 4.3.1: 
---
Version 4.3.1 only: 
PSI guides generated using the incorrect axis when planning with a perform monopost standard  baseplate and a BIO-RSA® Angled graft 
---
Version 4.2.2 & 4.3.1: 
Pyrocarbon humeral heads selectable with hrs flex procedures. Pyrocarbon heads are labeled for use with Tornier Flex only.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies

Tornier S.A.S.

161, Rue Lavoisier, Montbonnot-Saint-Martin, France, 38330

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-82972

Get notified

Receive emails about new and updated recall and safety alerts.