Health product recall

Bovie Medicals' J-Plasma Handpiece, Blade Tip, Pistol Grip (2018-02-08)

Starting date:
February 8, 2018
Posting date:
March 13, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-66122

Affected products

J-PLASMA HANDPIECE, BLADE TIP, PISTOL GRIP

Reason

Bovie Medical has become aware of an issue caused by incomplete insertion of the J-Plasma Handpiece cable plug into the generator receptacle. When the J-Plasma Handpiece with generator is used during surgical procedures without the cable plug being completely inserted into the generator, the green indicator light to the left of the receptacle will illuminate event with an incomplete cable plug insertion. There have been no patient injuries reported as a result of this issue.

Affected products

J-PLASMA HANDPIECE, BLADE TIP, PISTOL GRIP

Lot or serial number

All lots.

Model or catalog number

BVX-150B

BVX-150BPP

BVX-270B

BVX-330B

BVX-330BR

 

Companies
Manufacturer

Bovie Medical Corporation

5115 Ulmerton Road

Clearwater

33760

Florida

UNITED STATES