BS 3/8x3/32 L1.7 (Bubble Sensor for 3/8” X 3/32” tubing, length 1,7 m)
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| BS 3/8x3/32 L1.7 (Bubble Sensor for 3/8” X 3/32” tubing, length 1,7 m) | More than 10 numbers, contact manufacturer. | 70105.5720 |
Issue
Internal investigations have identified an issue with the durability of the connecting cable near the connection to the Bubble Sensor. Excessive bending of the cable can lead to poor contact, which may trigger errors on the connected medical device. These errors can occur temporarily when the cable is moved or permanently if the connection is fully compromised.
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies
Maquet Cardiopulmonary Gmbh
Kehler Strasse 31, Rastatt, Germany, 76437