CAPTUS 4000e (2020-03-02)
- Starting date:
- March 12, 2020
- Posting date:
- May 1, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-72919
Last updated: 2020-05-01
Affected Products
CAPTUS 4000e
Reason
Manufacturer received two notices of collimator detachment from Captus 4000 systems shipped in July and September of 2016. The collimator weighs 28 pounds and may cause injury if it falls on a patient or operator. Safety inspection and possible recall includes serial numbers 940103 through 940155.
Affected products
CAPTUS 4000e
Lot or serial number
- SN: 940001 through 940545
Model or catalog number
- 5430-30151
- 5430-30152
Companies
- Manufacturer
-
MIRION TECHNOLOGIES (CAPINTEC)
7 Vreeland Road
Florham PArk
07932
New Jersey
UNITED STATES