Health product recall

Cardinal Health Cardiac Electrophysiology (EP) Packs

Last updated

Summary

Product
Cardinal Health Cardiac Electrophysiology (EP) Packs
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare
Industry

Affected products

Affected products Lot or serial number Model or catalogue number
Cardiac Electrophysiology (EP) Packs 479584 SAN30EP'XXX'
Cardiac Electrophysiology (EP) Packs 473022 SAN30EP'XXX'

Issue

Cardinal Health was recently informed of a medical device recall from Medline Industries. Concerning the Namic Angiographic Control Syringes referenced. The recall is being issued due to a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold. 

Presource has completed its investigation regarding the recall and determined that customers may have received Presource kits containing the affected Medline industries component(s) and lot number(s).

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies

Cardinal Health 200, LLC Also Trading As Cardinal Health

3651 Birchwood Drive, Waukegan, Illinois, United States, 60085

Published by
Health Canada
Audience
Healthcare
Industry
Recall class
Type I
Recall date
Identification number
RA-81803