CARDIOSAVE® Hybrid And CARDIOSAVE® Rescue Intra-Aortic Balloon Pumps
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
CARDIOSAVE® Hybrid And CARDIOSAVE® Rescue Intra-Aortic Balloon Pumps |
More than 10 numbers, contact manufacturer. |
0998-00-0800-53 |
CARDIOSAVE® Hybrid And CARDIOSAVE® Rescue Intra-Aortic Balloon Pumps |
More than 10 numbers, contact manufacturer. |
0998-00-0800-65 |
Issue
Getinge has identified that some CARDIOSAVE® devices may contain a power management board (PMB) and/or solenoid control board (SCB) with one or more capacitors installed with reversed polarity. These capacitors may degrade over time, potentially causing CARDIOSAVE® device battery charging failures, an inability to power on, interruption of therapy, or unexpected shutdown.
The device may continue to function as intended, since the component degradation is not an immediate effect. Continued operation in this condition may eventually lead to a circuit malfunction, affecting critical system functions as outlined in the factors contributing to risk. Functional testing of the PMBs and SCBs may not identify every affected assembly unless capacitor polarity is specifically inspected and verified.
Additional information
Details
Datascope Corp.
1300 Macarthur Blvd., Mahwah, New Jersey, United States, 07430