Health product recall

CARDIOSAVE® Hybrid And CARDIOSAVE® Rescue Intra-Aortic Balloon Pumps

Brand(s)
Last updated

Summary

Product
CARDIOSAVE® Hybrid And CARDIOSAVE® Rescue Intra-Aortic Balloon Pumps
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

CARDIOSAVE® Hybrid And CARDIOSAVE® Rescue Intra-Aortic Balloon Pumps

More than 10 numbers, contact manufacturer.

0998-00-0800-53

CARDIOSAVE® Hybrid And CARDIOSAVE® Rescue Intra-Aortic Balloon Pumps

More than 10 numbers, contact manufacturer.

0998-00-0800-65

Issue

Getinge has identified that some CARDIOSAVE® devices may contain a power management board (PMB) and/or solenoid control board (SCB) with one or more capacitors installed with reversed polarity. These capacitors may degrade over time, potentially causing CARDIOSAVE® device battery charging failures, an inability to power on, interruption of therapy, or unexpected shutdown.

The device may continue to function as intended, since the component degradation is not an immediate effect. Continued operation in this condition may eventually lead to a circuit malfunction, affecting critical system functions as outlined in the factors contributing to risk. Functional testing of the PMBs and SCBs may not identify every affected assembly unless capacitor polarity is specifically inspected and verified.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies

Datascope Corp.

1300 Macarthur Blvd., Mahwah, New Jersey, United States, 07430

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-82982

Get notified