CARELINK SMARTSYNC DEVICE MANAGER (2020-12-18)
- Starting date:
- December 18, 2020
- Posting date:
- January 7, 2021
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-74709
Last updated:
2021-01-07
Affected Products
CARELINK SMARTSYNC DEVICE MANAGER
Reason
This update addresses a rare communication sequence during the first device interrogation with a Smartsync device manager that may result in the temporary suspension of some device features (i.e., battery measurements, capture management, atrial lead position check, effectivcrt algorithms, and adaptivcrt).This rare interaction results in temporary suspension of automatic threshold testing and output adjustments, and suspension of auto-optimization of CRT therapy. The issue is unlikely to result in clinical impact to the patient, and features are restored upon next programmer device interrogation or presence of a magnet.
Affected products
CARELINK SMARTSYNC DEVICE MANAGER
Lot or serial number
All lots.
Model or catalog number
24970A
Companies
- Manufacturer
-
Medtronic Inc.
710 Medtronic Parkway
Minneapolis
55432
Minnesota
UNITED STATES