Health product recall

CARESCAPE Monitoring System

Last updated

Summary

Product
CARESCAPE Monitoring System
Issue
Medical devices - Performance issue
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

CARESCAPE B650

More than 10 numbers, contact manufacturer.

MBB323
MBB313

CARESCAPE Canvas™ Smart Display

More than 10 numbers, contact manufacturer.

5897060

CARESCAPE Monitor B850

More than 10 numbers, contact manufacturer.

2095802-001
MBC313
MBC323

CARESCAPE Canvas™

More than 10 numbers, contact manufacturer.

5876727

Issue

Potential for loss of monitoring if, following the replacement of the CPU timekeeper battery, a mains power loss occurs during active monitoring.

Recall start date: April 22, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies

GE Healthcare Finland Oy

Kuortaneenkatu 2, Helsinki, Finland, 00510

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75486

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe