CellMek SPS
Brand(s)
Last updated
Summary
Product
CellMek SPS
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| CellMek SPS | SW Version: 2.3.90.0 | C44603 |
Issue
Beckman Coulter has confirmed an error in SW version 2.3.90.0 that causes samples to be dispensed unexpectedly close to the bottom of the output tubes. The intended design is for samples to be dispensed above the fluid surface into the output tubes.
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Haematology
Companies
| Beckman Coulter, Inc. |
| 250 S. Kraemer Blvd., Brea, California, United States, 92821 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-81436