Health product recall

Centricity High Acuity Anesthesia

Last updated

Summary

Product
Centricity High Acuity Anesthesia
Issue
Medical devices - Performance issue
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Centricity High Acuity Anesthesia

JGHCHAA

K1900AN

Issue

Patient information can be mixed when the same social security number "(ssn) or other patient identifier" is entered for two different patients and is sent from the system interface to centricity high acuity anesthesia and centricity high acuity critical care.

Recall start date: September 22, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies

Ge Healthcare Iits Usa Corp Also Doing Business As Ge Healthcare

500 W. Monroe Street, Chicago, Illinois, United States, 60661

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-64593

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe