Centricity High Acuity (CHA) System
Last updated
Summary
Product
Centricity High Acuity (CHA) System
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
---|---|---|
Centricity High Acuity Anesthesia |
JGHCHAA |
5854438-008 |
Centricity Anesthesia |
CAN1-C-Anesthesia |
M88021FD |
Issue
Centricity High Acuity (CHA) system discontinues infusion orders with a 'continuous' timing schedule once the ordered amount is given even if the prescribed time duration is not met.
Recall start date: April 22, 2025
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
Ge Healthcare Iits Usa Corp Also Doing Business As Ge Healthcare |
500 W. Monroe Street, Chicago, Illinois, United States, 60661 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77367
Get notified