Cetrimide Agar (CET) (2019-10-28)
- Starting date:
- October 26, 2019
- Posting date:
- January 6, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-71975
Last updated: 2020-01-07
Affected Products
Cetrimide Agar (CET)
Reason
Fertility issue on cetrimide agar. All impacted lots manufactured with the same semi finished product formulation, Biomerieux observed a low growth of pseudomonas aeruginosa ATCC 10145 (used in clinical bacteriology) and no (for one plate) or low growth for pseudomonas aeruginosa ATCC 9027 (used in industry bacteriology).
Affected products
Cetrimide Agar (CET)
Lot or serial number
1007392590
1007475130
1007512190
1007530990
Model or catalog number
43565
Companies
- Manufacturer
-
BioMérieux S.A
376 Chemin De L'orme
Marcy-l'etoile
69280
FRANCE