Cobas e801 Syringes (2018-03-23)
- Starting date:
- March 23, 2018
- Posting date:
- April 24, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-66520
Affected products
Cobas e801 Syringes
Reason
The plunger of the Cobas e801 syringe assembly may be mounted in a tilted position which applies mechanical stress to the plunger and may cause r1, r2 or prewash liquids to be pipetted incorrectly.
Affected products
Cobas e801 Syringes
Lot or serial number
- 1601-01 To 18E6-10
Model or catalog number
- 07682913001
Companies
- Manufacturer
-
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY