Combi Suite Angio Transfer Board (2018-11-29)
- Starting date:
- November 29, 2018
- Posting date:
- January 8, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-68680
Affected products
Combi Suite Angio Transfer Board
Reason
The springs, screws and threaded inserts in both handles of the transfer boards are magnetic and can cause signal loss in the region of knee and foot. The artifacts will be clearly recognizable. The impacted handles will be exchanged.
Affected products
Combi Suite Angio Transfer Board
Lot or serial number
- 1005
Model or catalog number
- 10849629
Companies
- Manufacturer
-
Siemens Healthcare GmbH
Henkestr. 127
Erlangen
91052
GERMANY