Health product recall

Corin Operating Tables

Brand(s)
Last updated

Summary

Product
Corin Operating Tables
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalogue number
Corin Operating Tables 7 770001F0
Corin Operating Tables 9 770001F0
Corin Operating Tables 13 770001F0
Corin Operating Tables 11 770001F0

Issue

Under certain conditions - such as the simultaneous use of the operating table with other medical devices that may cause interference, including High-Frequency (HF) surgical equipment - we have identified a potential issue that could prevent the device from performing as intended. If the issue occurs, the Corin operating table does not respond to any command. It is impossible to actuate any movement of the Corin operating table, neither with the universal remote control nor the override panel. We are not aware of instances where a patient was injured due to the described issue.

Recall start date: July 18, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies
Maquet Gmbh
Kehler Strasse 31, Rastatt, Germany, 76437
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77760

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe