Coulter DxH Diluent
Brand(s)
Last updated
Summary
Product
Coulter DxH Diluent
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Affected products
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| Coulter DxH Diluent | More than 10 numbers, contact manufacturer. | 628017 |
Issue
Beckman Coulter has confirmed that specific diluent lot numbers of Coulter DxH Diluent contribute to elevated platelet (PLT) Daily Checks background counts cycle when used with the UniCel DxH 600/690T/800/900 analyzers. This may result in elevated PLT background counts in Daily Checks ranging from 3.0 to 9.1 x 103 cells/μL.
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices
Companies
Beckman Coulter, Inc.
250 S. Kraemer Blvd., Brea, California, United States, 92821
Published by
Health Canada
Audience
General public
Healthcare
Industry
Recall class
Type II
Recall date
Identification number
RA-81285
Get notified
Receive emails about new and updated recall and safety alerts.